TopPette Adjustable Volume Mechanical Pipettes Volume Range 20-200μL | Each (7010101009)Price on request
SKU: 7010101009Brand: DLab Scientific

TopPette Adjustable Volume Mechanical Pipettes Volume Range 20-200μL | Each (7010101009)

Key features

  • 20-200 μL range — most commonly used pipette volume in molecular biology
  • Ergonomic lightweight body with low pipetting force
  • Digital display; calibration adjusted with the DLab calibration tool
  • Yellow tip cone fits universal 200 μL tips
  • Supplied with individual test certificate
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

The DLab TopPette 20-200 μL is a single-channel adjustable mechanical air-displacement pipette covering the most-used range in molecular biology. It features ergonomic design, digital display, calibration with the DLab calibration tool, and ISO 8655 compliant performance with individual test certificate.

Specifications

ModelTopPette (Cat. No. 7010101009)
BrandDLab Scientific (Beijing)
TypeSingle-channel, adjustable volume, air displacement
Volume Range20-200 μL
Increment (Step)1 μL (fine adjustment)
Accuracy (±%)±3.0% at 20 μL; ±1.0% at 100 μL; ±0.8% at 200 μL
Precision (CV%)≤1.5% at 20 μL; ≤0.4% at 100 μL; ≤0.3% at 200 μL
Tip Cone ColorYellow (compatible with 200 μL tips)
DisplayDigital mechanical counter
CalibrationAdjusted with the DLab calibration tool; refer to ISO 8655
AutoclavabilityLower shaft only (121°C, 20 min)
Weight~80 g
Length180 mm
StandardISO 8655; CE-marked

Applications

Used in molecular biology (PCR setup, DNA/RNA prep), immunoassays (ELISA), enzyme assays, cell culture media dispensing, and routine analytical chemistry. The 20-200 μL range is the workhorse for microplate-based workflows.

Setup guide

  1. Set volume via thumb wheel; verify on display.
  2. Attach tip firmly; pre-wet by aspirating/dispensing 3 times.
  3. Press plunger to first stop; immerse tip just below the surface, no deeper than 2 mm in the liquid.
  4. Slowly release plunger to aspirate; wait 1 second.
  5. Dispense to first stop, then second stop to blow out.

Compatibility

Compatible with universal 200 μL pipette tips. Optional: replacement seal kit; calibration tool; wall stand.

Precautions

Never pipette without a tip. Do not invert pipette with liquid in tip. Pre-wet tip 3 times for accurate delivery. Avoid pipetting concentrated acids/bases — use positive-displacement pipette instead.

Documents

FAQ

Can the TopPette 20-200 µL pipette be autoclaved?

DLab's pipette guide classes TopPette pipettes as semi-autoclavable: only the lower section can be autoclaved, at 121 C and 1 bar for 20 minutes. Let it cool and dry for 12 hours before use and check its performance afterwards. MicroPette Plus is the fully autoclavable DLab series.

How do I pipette correctly with the TopPette 20-200 µL?

Set the volume within the range only, approaching it from above, as forcing the knob past the limit can jam the mechanism. Pre-wet a tight-fitting tip 2-3 times. DLab's guide gives a maximum immersion depth of 2 mm for this volume class, so keep the tip just below the meniscus and hold the pipette vertical while aspirating.

How do I check or adjust the calibration of the TopPette 20-200 µL pipette?

To check volume, weigh dispensed distilled water on an analytical balance and convert the mass to volume with the ISO 8655-6 Z factor for the water temperature and pressure; 1 mL does not weigh exactly 1 g. To adjust, DLab's guide says to turn the calibration adjustment lock with the calibration tool. DLab states its specifications are guidelines and user calibration should refer to ISO 8655.

How should I store the 20-200 µL pipette and check it for leaks?

After use set the volume to the top of the range so the spring relaxes, and keep the pipette upright on a hook or pipette rack. Never lay it down with liquid in the tip. For a leak check, aspirate liquid, hold the pipette vertical for a few seconds and see if the level drops; if it does, check the tip fit, the spring and the piston.

When should I use reverse pipetting with the 20-200 µL pipette?

Use forward pipetting for aqueous, low-density liquids. DLab recommends reverse pipetting for viscous or volatile liquids: press to the second stop, release to aspirate more than the set volume, dispense to the first stop and discard the excess left in the tip. Pre-wet the tip by pipetting the liquid 2-3 times when you change volume or tip.