Hipette Single-channel Adjustable Volume color pipette 30-300ul | Each (7030601040)Price on request
SKU: 7030601040Brand: DLab Scientific

Hipette Single-channel Adjustable Volume color pipette 30-300ul | Each (7030601040)

Key features

  • 30-300 μL narrow-range pipette for higher precision in mid-volume applications
  • Color-coded body for quick identification
  • Ergonomic lightweight design; digital display; calibration adjusted with the DLab calibration tool
  • Universal blue tip cone fits 1000 μL tips
  • ISO 8655 compliant with individual test certificate
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

The DLab Hipette 30-300 μL is a color-coded single-channel adjustable mechanical pipette covering a narrow mid-volume range. The reduced range gives better accuracy vs full-range 100-1000 μL pipettes in the 30-300 μL region. Color-coded for quick identification in multi-pipette labs.

Specifications

ModelHipette (Cat. No. 7030601040)
BrandDLab Scientific (Beijing)
TypeSingle-channel, adjustable volume, air displacement, color-coded
Volume Range30-300 μL (narrow range)
Increment (Step)1 μL (fine adjustment)
Accuracy (±%)±1.5% at 30 μL; ±0.8% at 150 μL; ±0.8% at 300 μL
Precision (CV%)≤0.6% at 30 μL; ≤0.3% at 150 μL; ≤0.3% at 300 μL
Tip Cone ColorBlue
DisplayDigital mechanical counter
CalibrationAdjusted with the DLab calibration tool; refer to ISO 8655
AutoclavabilityLower shaft only (121°C)
Weight~90 g
Length200 mm
StandardISO 8655; CE-marked

Applications

Used in clinical chemistry, microbiology, immunoassay, and analytical labs requiring precise 30-300 μL dispensing. The narrow range is preferred for repeated dispensing of fixed volumes in QC protocols.

Setup guide

  1. Set volume; attach tip; pre-wet 3 times.
  2. Press to first stop; immerse tip just below the surface (see the pipette user guide for the maximum depth).
  3. Slowly release plunger; wait 1 second.
  4. Dispense to first stop then second stop.
  5. Eject tip into biohazard container.

Compatibility

Compatible with universal 1000 μL pipette tips. Optional: replacement seal kit; calibration tool; wall stand.

Precautions

Never pipette without a tip. Do not invert pipette with liquid in tip. Pre-wet tip 3 times for accurate delivery. Use only within calibrated range — do not force below 30 μL.

Documents

FAQ

What is the best way to pipette viscous liquids with the 30-300 µL Hipette?

DLab recommends reverse pipetting for viscous or volatile liquids: press to the second stop, release slowly to aspirate more than the set volume, dispense to the first stop and discard the leftover in the tip. Aspirate slowly to avoid bubbles and pre-wet the tip.

How do I check or adjust the calibration of the Hipette 30-300 µL pipette?

To check volume, weigh dispensed distilled water on an analytical balance and convert the mass to volume with the ISO 8655-6 Z factor for the water temperature and pressure; 1 mL does not weigh exactly 1 g. To adjust, DLab's guide says to turn the calibration adjustment lock with the calibration tool. DLab states its specifications are guidelines and user calibration should refer to ISO 8655.

How deep should the tip go when aspirating with the 30-300 µL Hipette?

Keep the tip just below the meniscus and hold the pipette vertical. Press smoothly to the first stop, release slowly and pause briefly before withdrawing the tip. DLab's guide lists maximum immersion depths of 2 mm for 20-200 uL and 3 mm for 100-1000 uL ranges, so keep it shallow.

How should I store the 30-300 µL pipette and check it for leaks?

After use set the volume to the top of the range so the spring relaxes, and keep the pipette upright on a hook or pipette rack. Never lay it down with liquid in the tip. For a leak check, aspirate liquid, hold the pipette vertical for a few seconds and see if the level drops; if it does, check the tip fit, the spring and the piston.

What causes liquid droplets or low volume with the 30-300 µL pipette?

DLab's troubleshooting table lists an unsuitable tip (use DLab original tips), poor tip wetting (repeat pipetting 4-5 times), an improperly seated tip, liquid dried in the tip cone (clean with alcohol), a dirty piston or a damaged O-ring. If the plunger sticks, clean and grease the piston and O-ring; other faults go back to the service centre.