Na Sodium Standard solution 150ppm, 6 x 14 ml | Each (3200457724)
Key features
- HORIBA Y-series Sodium (Na⁺) standard solution at 150 ppm (mg/L)
- Supplied as 6 × 14 ml dropper bottles— recap after use and never return used solution to the bottle
- Used as the low calibration point for the matching LAQUAtwin ion meter
- Convenient 14 ml size fits directly onto the LAQUAtwin flat sensor for one-drop calibration
Overview
HORIBA Y022L Sodium (Na⁺) Standard Solution (150 ppm / mg/L) is supplied as a pack of six 14 ml dropper bottles. It is used as the low calibration point for the LAQUAtwin Na-11 ion pocket meter only. Recap each bottle after use and never return used solution to the bottle.
Specifications
| Article number | 3200457724 |
|---|---|
| HORIBA model | Y022L |
| Parameter | Sodium (Na⁺) |
| Concentration | 150 ppm (mg/L) |
| Pack | 6 × 14 ml dropper bottles |
| Compatibility | LAQUAtwin Na-11 |
| Storage | 5-30 °C, tightly closed, away from direct sunlight |
| Shelf life | See bottle label |
| Brand | HORIBA Advanced Techno, Co., Ltd. |
Applications
Used as the low calibration point for the LAQUAtwin Na-11 ion pocket meter. The 150 ppm and 2000 ppm standards together form the two-point calibration required before measuring Sodium (Na⁺) in plant sap, soil extracts, irrigation water, food, beverage or biological samples.
Setup guide
- Open one 14 ml bottle of the low (150 ppm) standard.
- Apply a few drops directly onto the LAQUAtwin flat sensor (or onto sampling sheet B for 0.05 ml volumes).
- Press the calibration key — the meter will auto-recognise the 150 ppm standard.
- After calibration, rinse the sensor thoroughly with DI water.
- Repeat with the complementary 2000 ppm high standard to complete the two-point calibration, then proceed with sample measurement.
Compatibility
HORIBA LAQUAtwin Na-11 pocket meter; HORIBA LAQUAtwin 4M Kit (1300083234); HORIBA LAQUAtwin sampling sheet B (Y046); HORIBA 60 ml rinse bottle.
Precautions
For calibration use only; do not ingest. Avoid contact with skin and eyes. Recap each bottle immediately after dispensing to prevent evaporation and contamination. Do not return used standard to the bottle. Dispose of expired standard according to local regulations.
FAQ
What is the Sodium 150 ppm standard used for on the LAQUAtwin Na-11?
It is the low point of the two-point calibration of the LAQUAtwin Na-11. The meter recognises the 150 ppm and 2000 ppm sodium standards automatically by default, so calibrate with both.
How do I calibrate the LAQUAtwin Na-11 with the 150 ppm sodium standard?
Open the sensor cover, rinse the flat sensor with DI or tap water, dry it with a tissue and cover the whole sensor with a few drops of the 150 ppm standard. Close the cover, press CAL and confirm the value shown, then repeat for the 2000 ppm standard. Rinsing the sensor with a little of the next standard first improves accuracy.
Why does the Na-11 need both 150 ppm and 2000 ppm sodium standards?
A two-point calibration fits the sensor slope across the working range. The two standards are about ten-fold apart, so they bracket typical samples. HORIBA advises calibrating with solutions close to the expected sample value, and the high point can be changed between 160 and 9990 ppm if you use a different standard.
Does the 150 ppm sodium standard work for low-sodium samples with the Na-11?
It is the default low point, about 6.5 mmol/L, and the Na-11 measures down to 2 ppm. HORIBA advises calibrating with standards close to the expected value, and the low calibration value can be changed if your samples are much lower. Rinse well between standard and sample.
How should I store and handle the 150 ppm sodium standard bottles?
Keep each bottle tightly closed after dispensing and store it away from direct sunlight and heat, following any storage and expiry information printed on the label. Apply only the drops you need and never return used solution to the bottle, so the standard left in the bottle stays uncontaminated.