Ca Calcium Standard solution 2000ppm, 6 x 14 ml | Each (3200457727)Price on request
SKU: 3200457727Brand: HORIBA

Ca Calcium Standard solution 2000ppm, 6 x 14 ml | Each (3200457727)

Key features

  • HORIBA Y-series Calcium (Ca²⁺) standard solution at 2000 ppm (mg/L)
  • Supplied as 6 × 14 ml dropper bottles— recap after use and never return used solution to the bottle
  • Used as the high calibration point for the matching LAQUAtwin ion meter
  • Convenient 14 ml size fits directly onto the LAQUAtwin flat sensor for one-drop calibration
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

HORIBA Y051H Calcium (Ca²⁺) Standard Solution (2000 ppm / mg/L) is supplied as a pack of six 14 ml dropper bottles. It is used as the high calibration point for the LAQUAtwin Ca-11 ion pocket meter only. Recap each bottle after use and never return used solution to the bottle.

Specifications

Article number3200457727
HORIBA modelY051H
ParameterCalcium (Ca²⁺)
Concentration2000 ppm (mg/L)
Pack6 × 14 ml dropper bottles
CompatibilityLAQUAtwin Ca-11
Storage5-30 °C, tightly closed, away from direct sunlight
Shelf lifeSee bottle label
BrandHORIBA Advanced Techno, Co., Ltd.

Applications

Used as the high calibration point for the LAQUAtwin Ca-11 ion pocket meter. The 150 ppm and 2000 ppm standards together form the two-point calibration required before measuring Calcium (Ca²⁺) in plant sap, soil extracts, irrigation water, food, beverage or biological samples.

Setup guide

  1. Open one 14 ml bottle of the high (2000 ppm) standard.
  2. Apply a few drops directly onto the LAQUAtwin flat sensor (or onto sampling sheet B for 0.05 ml volumes).
  3. Press the calibration key — the meter will auto-recognise the 2000 ppm standard.
  4. After calibration, rinse the sensor thoroughly with DI water.
  5. Repeat with the complementary 150 ppm low standard to complete the two-point calibration, then proceed with sample measurement.

Compatibility

HORIBA LAQUAtwin Ca-11 pocket meter; HORIBA LAQUAtwin 4M Kit (1300083234); HORIBA LAQUAtwin sampling sheet B (Y046); HORIBA 60 ml rinse bottle.

Precautions

For calibration use only; do not ingest. Avoid contact with skin and eyes. Recap each bottle immediately after dispensing to prevent evaporation and contamination. Do not return used standard to the bottle. Dispose of expired standard according to local regulations.

FAQ

What is the Calcium 2000 ppm standard used for on the LAQUAtwin Ca-11?

It is the high point of the two-point calibration of the LAQUAtwin Ca-11. The meter recognises the 2000 ppm and 150 ppm calcium standards automatically by default, so calibrate with both.

How do I calibrate the LAQUAtwin Ca-11 with the 2000 ppm calcium standard?

Open the sensor cover, rinse the flat sensor with DI or tap water, dry it with a tissue and cover the whole sensor with a few drops of the 2000 ppm standard. Close the cover, press CAL and confirm the value shown, then repeat for the 150 ppm standard. Rinsing the sensor with a little of the next standard first improves accuracy.

Why does the Ca-11 need both 150 ppm and 2000 ppm calcium standards?

A two-point calibration fits the sensor slope across the working range. The two standards are about ten-fold apart, so they bracket typical samples. HORIBA advises calibrating with solutions close to the expected sample value, and the high point can be changed between 160 and 9990 ppm if you use a different standard.

What pH range should samples be in when calibrating and measuring with the Ca-11?

HORIBA specifies a working pH of 4 to 12 for the Ca-11, tested at 10 to the minus 3 mol/L calcium. Iron(II), zinc(II) and iron(III) ions are listed as interfering with the calcium sensor, so heavily contaminated samples may give unreliable readings. The 2000 ppm standard is about 50 mmol/L calcium.

How should I store and handle the 2000 ppm calcium standard bottles?

Keep each bottle tightly closed after dispensing and store it away from direct sunlight and heat, following any storage and expiry information printed on the label. Apply only the drops you need and never return used solution to the bottle, so the standard left in the bottle stays uncontaminated.