Toppette ingle-channel Adjustable Volume 2-20μl (Metal Nozzle) | Each (7030141005)
Key features
- 2-20 μL volume range with metal nozzle for superior chemical resistance
- Metal nozzle resists abrasion from repeated tip mounting
- Ergonomic lightweight body; digital display; calibration adjusted with the DLab calibration tool
- Compatible with universal 200 μL tips
- ISO 8655 compliant with individual test certificate
Overview
The DLab TopPette 2-20 μL with metal nozzle is a single-channel adjustable-volume mechanical pipette for low-volume precision work. The metal (stainless steel) nozzle offers superior durability vs plastic nozzles, especially when tips are mounted repeatedly or when aggressive solvents are used.
Specifications
| Model | TopPette Metal Nozzle (Cat. No. 7030141005) |
|---|---|
| Brand | DLab Scientific (Beijing) |
| Type | Single-channel, adjustable volume, air displacement |
| Volume Range | 2-20 μL |
| Increment (Step) | 0.1 μL (fine adjustment) |
| Accuracy (±%) | ±2.5% at 2 μL; ±1.5% at 10 μL; ±1.0% at 20 μL |
| Precision (CV%) | ≤1.5% at 2 μL; ≤0.5% at 10 μL; ≤0.4% at 20 μL |
| Tip Cone Material | Stainless steel (metal nozzle) |
| Tip Cone Color | White / natural metal |
| Display | Digital mechanical counter |
| Calibration | Adjusted with the DLab calibration tool; refer to ISO 8655 |
| Autoclavability | Lower shaft only (121°C) |
| Weight | ~80 g |
| Length | 180 mm |
| Standard | ISO 8655; CE-marked |
Applications
Used in molecular biology (PCR setup, qPCR, DNA quantification), drug discovery, and analytical chemistry requiring low-volume (2-20 μL) precision. The metal nozzle is preferred for high-throughput environments with frequent tip changes.
Setup guide
- Set volume via thumb wheel; verify on display.
- Attach tip firmly to metal nozzle; pre-wet 3 times.
- Press plunger to first stop; immerse tip just below the surface, no deeper than 2 mm.
- Slowly release plunger; wait 1 second.
- Dispense to first stop, then second stop to blow out.
Compatibility
Compatible with universal 200 μL pipette tips. Optional: replacement seal kit; calibration tool; pipette stand.
Precautions
Never pipette without a tip. Do not invert pipette with liquid in tip. Always pre-wet tip 3 times. Metal nozzle may corrode with prolonged exposure to strong acids/bases — rinse after use.
Documents
FAQ
Can the TopPette 2-20 µL pipette be autoclaved?
DLab's pipette guide classes TopPette pipettes as semi-autoclavable: only the lower section can be autoclaved, at 121 C and 1 bar for 20 minutes. Let it cool and dry for 12 hours before use and check its performance afterwards. MicroPette Plus is the fully autoclavable DLab series.
How do I pipette correctly with the TopPette metal nozzle 2-20 µL?
Fit the tip firmly on the nozzle, set the volume within range approaching from above, and pre-wet the tip 2-3 times. DLab's guide gives a maximum immersion depth of 2 mm for this volume class, so keep the tip just below the meniscus and hold the pipette vertical while aspirating.
How do I check or adjust the calibration of the TopPette 2-20 µL pipette?
To check volume, weigh dispensed distilled water on an analytical balance and convert the mass to volume with the ISO 8655-6 Z factor for the water temperature and pressure; 1 mL does not weigh exactly 1 g. To adjust, DLab's guide says to turn the calibration adjustment lock with the calibration tool. DLab states its specifications are guidelines and user calibration should refer to ISO 8655.
How should I store the 2-20 µL pipette and check it for leaks?
After use set the volume to the top of the range so the spring relaxes, and keep the pipette upright on a hook or pipette rack. Never lay it down with liquid in the tip. For a leak check, aspirate liquid, hold the pipette vertical for a few seconds and see if the level drops; if it does, check the tip fit, the spring and the piston.
When should I use reverse pipetting with the 2-20 µL pipette?
Use forward pipetting for aqueous, low-density liquids. DLab recommends reverse pipetting for viscous or volatile liquids: press to the second stop, release to aspirate more than the set volume, dispense to the first stop and discard the excess left in the tip. Pre-wet the tip by pipetting the liquid 2-3 times when you change volume or tip.