PAL-SALT Mohr digital salt meter, 0.00-10.0 % (g/100mL) (4251)Price on request
SKU: 4251Brand: ATAGO

PAL-SALT Mohr digital salt meter, 0.00-10.0 % (g/100mL) (4251)

Key features

  • Readings comparable to Mohr's titration method without reagents
  • Range 0.00-10.0 % (g/100mL), 0.01 % resolution up to 2.99 %
  • Accuracy ±0.05 % at 0.00-0.99 %, relative ±5 % at 1.00-10.0 %
  • Offset function aligns readings with known standards or applies a dilution factor
  • Durable titanium electrode and chemical-resistant PBT body
  • IP65 water protection, washable under running water
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

The ATAGO PAL-SALT Mohr is a pocket digital salt meter that uses the conductivity method to give readings similar to those obtained by Mohr's titration. It measures 0.00 to 10.0 % salt (g/100mL) with an offset function to compensate readings against known standards. Like other ATAGO PAL salt meters it uses a titanium electrode in a water-resistant PBT body.

Specifications

Product no4251
Model noPAL-SALT Mohr
ScaleSalt concentration (g/100mL), temperature
MethodConductivity
Measurement range0.00 to 10.0 % (g/100mL)
Resolution (0.00 to 2.99 %)0.01 %
Resolution (3.0 to 10.0 %)0.1 %
Accuracy (0.00 to 0.99 %)±0.05 %
Accuracy (1.00 to 10.0 %)Relative precision ±5 %
Offset functionYes
ElectrodeTitanium
Body materialPBT
Protection classIP65

Applications

Salt checks of soups, sauces, dressings, pickling brines, meat, fish and other foods where results must match Mohr titration values; solids are measured after a 1:10 dilution.

Setup guide

  1. Clean the sample stage and press ZERO with nothing on it (zero-set to air) at the start of each day.
  2. Dilute thick or concentrated samples, or anything above 10 % salt, so the reading falls within range.
  3. Apply 2-3 drops of sample to the sample stage.
  4. Press START and read the value after about 3 seconds.
  5. For hot or cold samples, leave the sample on the sensor about 30 seconds or repeat until readings stabilise.
  6. Rinse the unit under running water and wipe dry after use.

Precautions

Use distilled or tap water for dilutions, never mineral water. If readings become irregular, clean the stage with water or ethyl alcohol and, if needed, reference-set with a 2.50 % NaCl standard. Store the unit in its storage case when not in use for long periods.

FAQ

How is the PAL-SALT Mohr different from a standard PAL-SALT?

The PAL-SALT Mohr is tuned to give readings close to Mohr silver nitrate titration, expressed as g/100 mL, without chemicals. ATAGO notes that conductivity meters and Mohr titration can still give slightly different values for the same sample because the principles and units differ.

How do I measure solid foods with the PAL-SALT Mohr?

Mince or grind the food, mix it with water at a 1:10 ratio by weight, for example 10 g of sample with 90 g of water, and measure the liquid. Multiply the reading by 10, or set the offset function to the dilution factor so the display shows the original sample value.

How often should I zero the PAL-SALT Mohr and how do I clean it?

Zero-set with a clean, dry sample stage at the start of each day. After use rinse the unit under running water and wipe it dry. If readings become irregular, clean the stage with water or ethyl alcohol and, if needed, reference-set with a 2.50 % NaCl standard.

What range does the PAL-SALT Mohr cover and what if my sample is saltier?

It measures 0.00 to 10.0 % salt (g/100 mL), with 0.01 % resolution up to 2.99 %. Samples above 10 %, or thick concentrated samples, should be diluted with distilled or tap water (not mineral water) so the reading falls within range, then multiplied by the dilution factor.

How should I measure hot or cold samples with the PAL-SALT Mohr?

Conductivity depends on temperature, so let the sample sit on the sensor for about 30 seconds, or repeat the measurement until readings stabilise, before recording the value. Keep the unit in its storage case when not in use for long periods.