LAQUAtwin 4EC-33 conductivity/TDS pocket meter (3201077524)
Key features
- Measures conductivity and TDS directly from a 0.12 ml sample, no beaker needed
- Conductivity auto range 0 µS/cm to 199.9 mS/cm; TDS 0.0 to 9990 ppm
- 4 TDS conversion curves: adjustable factor 0.4 to 1.0, EN 27888, 442 and NaCl
- Up to 4-point auto calibration with auto recognition of 84 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm
- ±2% full scale to 19.99 mS/cm, ±5% full scale from 20.0 to 199.9 mS/cm
- IP67 waterproof with temperature display 0 to 50.0 °C and about 400 hours battery life
Overview
The LAQUAtwin 4EC-33 is the top model of HORIBA's 4-electrode 4EC pocket series, measuring conductivity, total dissolved solids (TDS) and temperature. It reads from as little as 0.12 ml placed on the sensor and offers four selectable TDS curves for different sample types. The meter is supplied in a carrying case with 1413 µS/cm and 12.88 mS/cm standards and conditioning solution.
Specifications
| Product no | 3201077524 |
|---|---|
| Model no | LAQUAtwin 4EC-33 |
| Measurement principle | 4-electrode method |
| Minimum sample volume | 0.12 ml |
| Conductivity range | 0 µS/cm to 199.9 mS/cm (4 auto ranges) |
| TDS range 1 | 0.0 to 99.9 ppm (0.1 ppm resolution) |
| TDS range 2 | 100 to 999 ppm (1 ppm resolution) |
| TDS range 3 | 1000 to 9990 ppm (10 ppm resolution) |
| TDS curves | Factor 0.4 to 1.0 (default 0.5), EN 27888, 442, NaCl |
| Accuracy | ±2% FS (0 to 19.99 mS/cm), ±5% FS (20.0 to 199.9 mS/cm) |
| Calibration points | Up to 4 |
| Auto standard recognition | 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm |
| Temperature display | 0 to 50.0 °C |
| Temperature adjustment | 5 to 40.0 °C |
| Temperature compensation | Automatic, 2%/°C fixed |
| Display | Custom monochrome LCD with backlight |
| Operating conditions | 5 to 40 °C, 85% RH or less (no condensation) |
| Protection | IP67 waterproof and dustproof |
| Power | 2 x CR2032 batteries (not included) |
| Battery life | Approx. 400 hours continuous (backlight off) |
| Material | ABS epoxy body, carbon sensor |
| Dimensions | 164 x 29 x 20 mm (excluding projections) |
| Weight | Approx. 45 g (meter only, without batteries) |
| Replacement sensor | S073-4EC |
| Warranty | 2 years meter, 6 months sensor |
Applications
Conductivity and TDS testing of drinking, process and irrigation water, hydroponic and fertigation nutrient solutions, aquariums and small-volume lab samples.
Setup guide
- Fit 2 x CR2032 batteries (not included) and slide the sensor onto the meter, checking the waterproof gasket is clean and seated.
- Condition the new sensor: place a few drops of the supplied conditioning solution on the sensor, wait about 20 minutes, then rinse with pure water.
- Calibrate by placing a few drops of 1413 µS/cm standard on the sensor and pressing CAL; the meter recognises the standard automatically.
- Repeat with 12.88 mS/cm standard for a second point, rinsing between standards.
- Select the TDS curve (factor, EN 27888, 442 or NaCl) to suit the sample, then rinse and blot the sensor.
- Place a few drops of sample on the sensor, close the cover and read conductivity or TDS when the stability indicator appears; rinse with pure water after use.
In the box
- LAQUAtwin 4EC-33 meter with replaceable 4-cell conductivity sensor
- 1413 µS/cm conductivity standard (14 ml)
- 88 mS/cm conductivity standard (14 ml)
- Conditioning solution (4 ml)
- Dropper
- Instruction manual and quick manual
- Carrying case
Precautions
Do not measure organic solvents, oils, adhesives, cement, alcohols, surfactants, concentrated acids (pH 0-2) or concentrated alkalis (pH 12-14), as these can damage the sensor; rinse with pure water straight away after any contact. The meter is IP67 only when the sensor is properly fitted, and it uses CR2032 coin cells (not included), which are dangerous if swallowed, so keep batteries away from children. Sensors of the older LAQUAtwin EC-11/22/33 are not compatible with the 4EC series; order S073-4EC.
FAQ
What TDS curves does the LAQUAtwin 4EC-33 offer and when should I use each?
It offers four TDS conversions: an adjustable TDS factor (0.4 to 1.0, default 0.5, for general use based on KCl), NaCl for aquaculture and pickling, 442 for boiler water and HVAC, and EN 27888 for environmental water. Pick the curve that matches the dissolved salts in your sample.
What TDS and conductivity ranges does the 4EC-33 measure?
Conductivity covers 0 uS/cm to 199.9 mS/cm in four auto ranges. TDS reads 0.0 to 99.9 ppm in 0.1 ppm steps, 100 to 999 ppm in 1 ppm steps and 1000 to 9990 ppm in 10 ppm steps. Accuracy is plus or minus 2 percent of full scale up to 19.99 mS/cm and 5 percent above.
How do I calibrate the 4EC-33?
It supports up to 4 calibration points and recognises 84 uS/cm, 1413 uS/cm, 12.88 mS/cm and 111.8 mS/cm standards automatically; the kit supplies the 1413 uS/cm and 12.88 mS/cm standards. Apply a few drops to the flat sensor, press CAL and rinse with pure water between standards.
Can I use my old EC-33 sensor on the LAQUAtwin 4EC-33?
No. HORIBA states the sensors for the EC series and the 4EC series are not compatible. The 4EC-33 needs the S073-4EC 4-cell sensor (part 3201079162), whereas the EC-33 uses S070. Check that the sensor and meter are both printed COND 4-cell before connecting.
How much sample does the 4EC-33 need and how is it powered?
A sample of 0.12 mL placed on the flat sensor is enough for conductivity, TDS and temperature. The meter uses two CR2032 batteries, which are not included, and runs about 400 hours continuously with the backlight off. It is IP67 rated only when the sensor is fitted correctly.