Scichemtech
Categories33
Region · LanguageUZ · English
United Arab Emirates

English

Saudi Arabia

English

Qatar

English

Oman

English

Kuwait

English

Bahrain

English

Middle East & North Africa

English

Türkiye

TürkçeEnglish

Kazakhstan

РусскийҚазақшаEnglish

Uzbekistan

O'zbekchaРусскийEnglish

+971 50 391 6126WhatsAppsales@sctme.ae
LAQUA DO210 dissolved oxygen handheld meter kit (3200779530)Price on request
SKU: 3200779530Brand: HORIBA

LAQUA DO210 dissolved oxygen handheld meter kit (3200779530)

Key features

  • Measures dissolved oxygen 0.0 to 20.00 mg/L and 0.0 to 200.0% saturation
  • 01 mg/L or 0.1% resolution with ±0.1 mg/L accuracy
  • Compensates for salinity (0.0 to 40.0 ppt), temperature and barometric pressure
  • Kit includes 9552-20D galvanic DO probe with built-in temperature sensor and 2 m cable
  • Auto data logging with 500-reading memory
  • IP67 waterproof, shock-resistant housing with over 500 hours battery life
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

The LAQUA DO210-K kit combines HORIBA's DO210 waterproof handheld dissolved oxygen/temperature meter with a 9552-20D galvanic DO probe on a 2 m cable. It reads DO in mg/L and % saturation and corrects for salinity, temperature and atmospheric pressure. Up to 2-point calibration (air and optional zero) keeps readings reliable in the field or lab.

Specifications

Product no3200779530
Model noLAQUA DO210 (kit DO210-K)
ParametersDissolved oxygen, temperature
DO range (mg/L)0.0 to 20.00 mg/L
DO range (% saturation)0.0 to 200.0%
DO resolution0.01 mg/L, 0.1%
DO accuracy±0.1 mg/L
Salinity compensation0.0 to 40.0 ppt
Barometric pressure compensationYes
Temperature compensationYes
DO probe typeGalvanic with integrated temperature sensor
Calibration pointsUp to 2
Data memory500 readings, auto data log
Meter inputsBNC (DO electrode), phono (temperature sensor)
Probe supplied9552-20D galvanic DO probe, 2 m cable
Temperature-30.0 to 130.0 °C, 0.1 °C resolution, ±0.5 °C
Measurement modesAuto stable, auto hold, real time
Auto shut-offProgrammable, up to 30 minutes
DisplayCustom LCD with white LED backlight (50 x 50 mm)
HousingIP67, shock and scratch resistant, non-slip
Power2 x AA batteries (not included), over 500 h life
Dimensions160 x 80 x 40.6 mm
WeightApprox. 216 g (without batteries)
Warranty3 years (meter)

Applications

Dissolved oxygen monitoring in aquaculture, wastewater aeration, rivers and lakes, and lab samples.

Setup guide

  1. Insert 2 x AA batteries and connect the 9552-20D probe to the meter.
  2. Enter the sample salinity and barometric pressure in the setup menu if needed.
  3. Rinse the probe with clean water, blot dry so no droplets remain on the membrane or temperature sensor, and hold it still in clean air.
  4. Press CAL, then ENT to confirm the air calibration; optionally press CAL twice and calibrate in zero solution (freshly prepared sodium sulfite, 1 g/L or more).
  5. Rinse the probe, immerse it so the membrane and temperature sensor are fully covered, and read DO in mg/L or %.

In the box

  • LAQUA DO210 dissolved oxygen/temperature meter
  • 9552-20D galvanic DO probe with built-in temperature sensor, 2 m cable
  • Carrying case
  • Instruction manual and quick guide

Precautions

Perform air calibration in clean air away from sudden temperature changes, rain or direct wind, and make sure no water droplets remain on the membrane. Fully immerse the membrane and temperature sensor when measuring. Batteries are not included.

FAQ

How does the LAQUA DO210 measure dissolved oxygen and do I need to wait for polarisation?

The 9552-20D is a membrane galvanic probe with an integrated temperature sensor, covering 0 to 20.00 mg/L and 0 to 50 C. Galvanic electrodes need no external voltage and no warm-up time, so they are ready as soon as they are connected and calibrated, and their electrolyte can be used indefinitely.

How do I calibrate the DO210 and how often?

For most work an air calibration is enough: calibrate in clean air with no water droplets on the membrane or temperature sensor, after setting local barometric pressure. A second zero point can be added in oxygen-free solution (freshly prepared sodium sulfite, 1 g or more per litre) when low DO matters. HORIBA advises checking calibration daily and after temperature or pressure changes.

Does the DO210 compensate for salinity and pressure, and does the sample need stirring?

Yes. The meter compensates for temperature, barometric pressure and salinity from 0.0 to 40.0 ppt, which you enter for the sample. Keep the temperature sensor submerged, avoid air bubbles, and move the probe or stir the sample, because the probe consumes oxygen at the membrane and the reading would otherwise fall.

How do I care for and replace the 9552-20D DO tip?

The probe uses a replaceable DO tip (5402) with a very thin membrane, so keep it away from hard surfaces. Wipe a dirty membrane gently with a soft cloth and rinse with clean water. If calibration fails, replace the DO tip. The internal electrolyte is highly concentrated potassium hydroxide, so wash skin or eyes immediately with plenty of water after contact.

What is the difference between the LAQUA DO210 and DO220, and what is in the kit?

The DO210 stores 500 readings with no real-time clock. The DO220 stores 1000 readings, adds a real-time clock with date and time stamps and supports PC (USB) or printer (RS232) connection with optional cables. The DO210-K contains the meter, a 9552-20D galvanic probe on a 2 m cable, a carrying case and the manual.