SCT-PH-MINI pH mini controller, horizontal (SCT-108.005.01)
Key features
- Online pH measurement and control for industrial processes
- Simple and compact design
- Meets international standards
- Easy to install, operates safely with low maintenance
- Insulated 4-20 mA output
- Large, clear LCD display
Overview
The SCT-pH-MINI is a panel-mounted online pH controller for industrial measurement and control of pH. It measures 0.00 to 14.00 pH with a resolution of 0.02 pH and provides an insulated 4-20 mA output plus high and low contact outputs. The horizontal 48 x 96 mm housing fits a 45 x 91 mm panel opening.
Specifications
| Product no | SCT-108.005.01 |
|---|---|
| Model no | SCT-PH-MINI |
| Measurement range | 0.00~14.00 pH |
| Resolution ratio | 0.02 pH |
| Accuracy | ±0.1 pH |
| Display | 3½-digit apheliotropic liquid crystal display |
| Calibration mode | three-point calibration (automatic identification of solutions) |
| Manual temperature compensation | 0~80°C |
| Temperature of measured solutions | 0~80°C |
| Compression resistance of electrodes | 0~0.6 MPa |
| Measurement distance | standard configuration 10 m (≤40 m) |
| Output current | insulated 4~20 mA (≤500Ω) |
| Control output | maximum and minimum (constantly open) contact point output |
| Capacity of contact points | 3A/AC250V (with resistant loads) |
| Power supply | AC 220V~240V |
| Environmental conditions | Temp: 0 to 50°C, Humidity: ≤85% |
| External dimensions | 48×96×100 mm (height × width × depth) |
| Face opening | 45×91 mm |
| Installation method | panel-mounted |
| Ordering code | SCT-108.005.01 (SCT-pH-MINI) |
| Weight | 1 kg |
Applications
Used in swimming pool pH control, aquaculture water treatment, hydroponics nutrient management, industrial process water, and small-scale wastewater monitoring. The high and low contact outputs can switch acid/base dosing equipment, through an interposing relay where needed, to hold the setpoint pH.
Setup guide
- Mount controller in panel (cutout per dimensions); secure with clamps.
- Wire power (230V AC) and relay output to dosing pump (per wiring diagram).
- Connect pH sensor; immerse sensor in flow cell or inline tee.
- Calibrate: immerse in pH 7 buffer, press CAL; rinse; immerse in pH 4 buffer, press CAL.
- Set setpoint + hysteresis; enable relay; verify dosing pump actuates on out-of-range pH.
In the box
- SCT-pH-MINI controller
- pH probe with 10 m cable
- Fixation clamps
- Instruction manual
- Optional: SCT-108.005.01A pH probe with 10 meter cable
- Optional: SCT-108.002.AA pH 4.00 buffer solution - 500 mL
- Optional: SCT-108.002.AG pH 7.00 buffer solution - 500 mL
- Optional: SCT-108.002.AM pH 10.01 buffer solution - 500 mL
- Optional: SCT-108.002.CA electrode storage solution - 500 mL
- Optional: SCT-108.005.01B mounting clamps
Compatibility
Includes: pH controller, pH sensor (combination electrode), fixation clamps (2x), instruction manual. Compatible with: standard 230V dosing pumps; chart recorder.
Precautions
Always de-energize power before wiring to relay contacts. Keep pH sensor wetted at all times — drying out damages glass membrane. Calibrate weekly for accurate control. Install check valve on dosing line to prevent backflow.
Documents
FAQ
How do I calibrate the SCT-PH-MINI with three buffers?
The controller supports three-point calibration with automatic recognition of the buffer solutions. Use fresh pH 4.00, 7.00 and 10.01 buffers, rinse the electrode between solutions, and make sure the buffers bracket the pH you expect to measure. Buffer pH shifts with temperature, so calibrate at about the same temperature as the process.
How often should the SCT-PH-MINI pH electrode be calibrated, and when should it be replaced?
Calibrate a new electrode, after cleaning, and after long storage; weekly calibration is recommended for accurate control, and dirty or hot samples may need daily checks. Replace the electrode if it will not calibrate with fresh buffers, responds slowly or drifts after cleaning. Where a slope is shown, roughly 95 to 105 percent is generally treated as good, but the acceptable lower limit varies between makers, so check the SCT manual.
How should the SCT-PH-MINI pH electrode be stored and kept wet?
Never let the glass membrane dry out. When the line is shut down or the probe is removed, keep the electrode in electrode storage solution, which SCT lists as an accessory; pH 4 or pH 7 buffer is acceptable for short periods. Avoid storing it in distilled or deionised water unless the electrode maker says so. A dried electrode may recover after soaking in storage solution for several hours, otherwise replace it.
How do I connect the SCT-PH-MINI to a dosing pump or PLC?
The high and low contact outputs are rated 3 A at 250 V AC for resistive loads, so for a pump motor or large solenoid use an interposing relay or the pump's control input. The insulated 4-20 mA output drives up to 500 ohm loads for a PLC or recorder, where 4 mA is a live zero that lets a broken wire be told apart from a real low reading. Fit a check valve on the dosing line.
What panel cut-out does the horizontal SCT-PH-MINI need, and how long can the cable be?
The horizontal housing is 48 x 96 mm face by 100 mm deep and mounts in a 45 x 91 mm panel opening with the supplied fixation clamps. The electrode comes on a 10 m cable, and the specified maximum measuring distance is 40 m. The electrode is rated for 0 to 0.6 MPa and sample temperatures from 0 to 80 C.