514-20, Conductivity/Salinity conditioning solution 6 x 14 ml | 6 x 14 ml (3999960114)
Key features
- HORIBA 514-20 conditioning solution for conductivity/salinity flat sensors, supplied as 6 × 14 ml dropper bottles
- Conditions LAQUAtwin EC/Salt flat sensors and removes residues
- Restores response speed and accuracy after extended use or measurement of fouling samples
- Small-volume dropper bottles — apply a few drops directly onto the LAQUAtwin flat sensor
- Recommended as part of routine maintenance for LAQUAtwin EC-11/EC-22 and Salt-11 pocket meters
Overview
HORIBA 514-20 (article 3999960114) is a conditioning solution for the flat conductivity/salinity sensors used in the LAQUAtwin EC-11, EC-22 and Salt-11 pocket meters, supplied as a pack of six 14 ml dropper bottles. It is a surfactant-based solution that conditions the sensor and removes residues, restoring response speed and accuracy after extended use or measurement of fouling samples.
Specifications
| Article number | 3999960114 |
|---|---|
| HORIBA model | 514-20 |
| Product | Conditioning Solution for conductivity/salinity flat sensors |
| Pack | 6 × 14 ml dropper bottles |
| Compatibility | LAQUAtwin EC-11, EC-22, Salt-11 (and other HORIBA flat conductivity sensors) |
| Function | Conditions the sensor and removes residues (surfactant solution) |
| Storage | 5-30 °C, tightly closed, away from direct sunlight |
| Shelf life | See bottle label |
| Brand | HORIBA Advanced Techno, Co., Ltd. |
Applications
Used as a routine maintenance reagent for the flat conductivity/salinity sensors of HORIBA LAQUAtwin EC-11, EC-22 and Salt-11 pocket meters. Recommended when the sensor response becomes slow, calibration becomes difficult, or after measuring samples that leave deposits on the electrodes (e.g. food brines, soil extracts, viscous liquids).
Setup guide
- Remove the sensor from the LAQUAtwin meter (or expose the flat sensor face).
- Open one 14 ml bottle of 514-20 conditioning solution.
- Apply a few drops to cover the flat sensor surface and let it stand for 10-30 minutes to condition the sensor and remove residues.
- Rinse the sensor thoroughly with DI water and gently blot dry (do not wipe the sensor surface).
- Recalibrate with the appropriate HORIBA conductivity or salinity standard (e.g. 1,413 µS/cm or 0.50 % NaCl) before resuming sample measurement.
Compatibility
HORIBA LAQUAtwin EC-11; LAQUAtwin EC-22; LAQUAtwin Salt-11; and other HORIBA flat-sensor conductivity/salinity pocket meters. Not for use with pH or ion-selective (Na, K, NO3, Ca, F) sensors.
Precautions
For sensor conditioning use only; do not ingest. Avoid contact with skin and eyes — rinse with water if spilled. Do not use on pH or ion-selective sensors. Recap each bottle immediately after dispensing. Dispose of expired solution according to local regulations.
FAQ
What is HORIBA 514-20 conditioning solution used for?
It is a conditioning solution for HORIBA LAQUAtwin conductivity sensors. HORIBA's technical tip says to place a few drops on the sensor and leave it for 10 to 30 minutes, and the same procedure is used to remove residues.
When should I condition a LAQUAtwin conductivity sensor?
Condition it before first use or after dry storage, and use the same procedure to remove sample residues from the sensor. Calibrate with fresh conductivity standards afterwards.
Can I use 514-20 on pH or ion sensors?
No. It is intended for conductivity/salinity sensors only; do not use it on pH or ion-selective sensors.
What if I run out of 514-20 conditioning solution?
HORIBA's technical tip says diluted detergent (5 ml liquid detergent in 100 ml water) can be used. For stubborn deposits use a stronger cleaning solution, never organic solvents such as acetone or ethanol, which can damage the sensor.