TopPette Adjustable Volume Mechanical Pipettes 5-50μL | Each (7010101006)
Key features
- Lightweight, ergonomic design with low pipetting force — reduces repetitive strain injury
- Digital volume display clearly reads current setting; calibration adjusted with the DLab calibration tool
- Volume range 5-50 μL with fine volume adjustment via thumb wheel
- ISO 8655 compliant; each pipette supplied with individual test certificate
- Universal tip cone accepts most standard 200 μL tips
Overview
The DLab TopPette is a single-channel adjustable-volume mechanical air-displacement pipette covering the 5-50 μL range. It uses a digital volume display, ergonomic handle design, and a calibration adjustment lock turned with the DLab calibration tool. The pipette is calibrated per ISO 8655 and supplied with an individual test certificate.
Specifications
| Model | TopPette (Cat. No. 7010101006) |
|---|---|
| Brand | DLab Scientific (Beijing) |
| Type | Single-channel, adjustable volume, air displacement |
| Volume Range | 5-50 μL |
| Increment (Step) | 0.5 μL |
| Accuracy (±%) | ±2.00% at 5 μL; ±0.90% at 25 μL; ±0.60% at 50 μL |
| Precision (CV%) | ≤1.5% at 5 μL; ≤0.6% at 25 μL; ≤0.4% at 50 μL |
| Tip Cone Color | Yellow (compatible with standard 200 μL tips) |
| Display | Digital mechanical counter |
| Calibration | Adjusted with the DLab calibration tool; refer to ISO 8655 |
| Autoclavability | Lower shaft only (121°C, 1 bar, 20 min) |
| Weight | ~80 g |
| Length | 180 mm |
| Performance Standard | ISO 8655; CE-marked |
Applications
Used in molecular biology, biochemistry, clinical diagnostics, and analytical chemistry labs for precise liquid handling in PCR setup, ELISA, sample dilution, reagent addition, and microplate loading. Suitable for routine bench work requiring 5-50 μL volumes.
Setup guide
- Select volume via thumb wheel; verify on digital display.
- Attach compatible tip firmly; pre-wet by aspirating/dispensing 3 times.
- Press plunger to first stop; immerse tip just below the surface (see the pipette user guide for the maximum depth).
- Slowly release plunger to aspirate; wait 1 second for full fill.
- Dispense by pressing to first stop, then to second stop to blow out residual liquid.
Compatibility
Compatible with universal 200 μL pipette tips (DLab, Eppendorf, Axygen, Sartorius). Optional: replacement shaft/seal kit; calibration tool kit; wall mount holder.
Precautions
Never pipette without a tip attached — damage to piston. Do not invert or lay pipette on its side with liquid in the tip — liquid entering the shaft damages the piston. Always pre-wet the tip by aspirating and dispensing 3 times for accurate delivery. Do not exceed the maximum volume.
Documents
FAQ
Can the TopPette 5-50 µL pipette be autoclaved?
DLab's pipette guide classes TopPette pipettes as semi-autoclavable: only the lower section can be autoclaved, at 121 C and 1 bar for 20 minutes. Let it cool and dry for 12 hours before use and check its performance afterwards. MicroPette Plus is the fully autoclavable DLab series.
How do I pipette correctly with the TopPette 5-50 µL?
Set the volume within the range only, approaching it from above, as forcing the knob past the limit can jam the mechanism. Pre-wet a tight-fitting tip 2-3 times. Keep the tip just below the meniscus, hold the pipette vertical, press smoothly to the first stop and release slowly, because fast aspiration creates bubbles that block the tip.
How do I check or adjust the calibration of the TopPette 5-50 µL pipette?
To check volume, weigh dispensed distilled water on an analytical balance and convert the mass to volume with the ISO 8655-6 Z factor for the water temperature and pressure; 1 mL does not weigh exactly 1 g. To adjust, DLab's guide says to turn the calibration adjustment lock with the calibration tool. DLab states its specifications are guidelines and user calibration should refer to ISO 8655.
How should I store the 5-50 µL pipette and check it for leaks?
After use set the volume to the top of the range so the spring relaxes, and keep the pipette upright on a hook or pipette rack. Never lay it down with liquid in the tip. For a leak check, aspirate liquid, hold the pipette vertical for a few seconds and see if the level drops; if it does, check the tip fit, the spring and the piston.
When should I use reverse pipetting with the 5-50 µL pipette?
Use forward pipetting for aqueous, low-density liquids. DLab recommends reverse pipetting for viscous or volatile liquids: press to the second stop, release to aspirate more than the set volume, dispense to the first stop and discard the excess left in the tip. Pre-wet the tip by pipetting the liquid 2-3 times when you change volume or tip.