pH 7,00 buffer solution, 500 ml, accuracy ± 0,01 @ 25°C * | 500ml/ Bottle (1300045608)Price on request
SKU: 1300045608Brand: HORIBA

pH 7,00 buffer solution, 500 ml, accuracy ± 0,01 @ 25°C * | 500ml/ Bottle (1300045608)

Key features

  • NIST-traceable pH 7.00 (neutral) buffer standard, accuracy ±0.01 @ 25°C
  • Supplied in a 500 ml dosing bottle for convenient, drip-free dispensing
  • Comes with a certificate of analysis for full metrological traceability
  • Coloured green per HORIBA convention for instant visual identification as pH 7 (neutral) buffer
  • Serves as the mandatory zero-point / slope reference for any pH electrode calibration
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

HORIBA pH 7.00 (±0.01 at 25°C) buffer solution is a NIST-traceable neutral calibration standard supplied in a 500 ml dosing bottle with certificate of analysis. It is the mandatory reference point for any pH electrode calibration (the slope is established against pH 4 and/or pH 10).

Specifications

Article number1300045608
ProductpH 7.00 Buffer Solution
pH value @ 25°C7.00 ±0.01
Volume500 ml
PackagingDosing bottle with dropper cap
TraceabilityNIST-traceable, supplied with certificate
Temperature reference25 °C
ColourGreen (HORIBA colour convention)
Shelf life2 years sealed, expiry date on bottle label
Storage5-30 °C, tightly closed, away from direct sunlight
BrandHORIBA Advanced Techno, Co., Ltd.

Applications

Used as the mandatory neutral (zero-point) calibration standard for any HORIBA or third-party pH meter and electrode that uses USA or NIST pH buffer sets. Required for both single-point, two-point and three-point pH calibrations across laboratory, field, environmental, food, and water-quality applications.

Setup guide

  1. Dispense a few drops of fresh pH 7.00 buffer onto the electrode (or into a clean beaker).
  2. Immerse the pH electrode and start calibration.
  3. Wait for the reading to stabilise and confirm the pH 7.00 (neutral) calibration point — this is the mandatory zero-point.
  4. Rinse the electrode thoroughly with DI water before moving to additional buffers (pH 4.01 / pH 10.01).
  5. Discard used buffer — never return it to the bottle.

Compatibility

HORIBA LAQUAtwin pH-11/22/33 pocket meters; HORIBA LAQUA PH200/PH220/PH1100/PH1200/PH1300/PH1500/PH2000 series; HORIBA twin pH B-71X series; any pH meter that recognises USA or NIST pH buffer sets at pH 7.00.

Precautions

For calibration use only; do not ingest. Avoid contact with skin and eyes — rinse with water if spilled. Recap the bottle immediately after dispensing to prevent contamination and evaporation.

FAQ

Is the HORIBA 7.00 500 ml buffer traceable and does it come with a certificate?

HORIBA states that its LAQUA buffers and standard solutions in dosing bottles are traceable to NIST standard reference materials and that each solution is supplied with a certificate of analysis. The lot number on the bottle links the solution to its traceability record.

How long does the pH 7.00 buffer last, unopened and after opening?

HORIBA gives a 2 year shelf life for sealed LAQUA dosing bottles stored as instructed, with the expiry date on the label. After opening, only a small amount of air gets into the bottle, so usable life is extended. Recap immediately after use to limit air contact.

Why is this buffer supplied in a dosing bottle and can I pour leftover back?

The dosing bottle delivers solution through a side straw into a separate calibration chamber, so used solution never returns to the bottle. Never pour used buffer back, because contamination shifts the value and spoils later calibrations. Discard the solution left in the chamber after each calibration.

What is the pH 7.00 buffer used for and where do I find its temperature table?

It is the neutral zero point and the first point of any calibration; two-point calibration pairs it with 4.01 or 10.01 depending on the sample range. The nominal value applies at 25 C; the bottle label carries a temperature table, so check the value at your actual buffer temperature if the meter does not compensate automatically.

Which order should I calibrate with the pH 7.00 buffer?

Start with pH 7.00 to set the zero point, then add pH 4.01 for acidic samples or pH 10.01 for alkaline ones. Rinse the electrode with deionised water and blot it dry between buffers so nothing carries over.