LAQUAtwin SALT-11 salinity (NaCl) Pocket meter with plastic carry-case, 14 ml buffer/standard solutions, 2 x CR2032 batteries, dropper | Each (3200689158)
Key features
- Only pocket meter that directly measures salt concentration (as NaCl or seawater equivalent) from a 0.12 ml sample drop
- Two selectable salinity curves — NaCl and Seawater — for accurate conversion in different sample matrices
- Up to 2-point calibration with 0.50 % and 5.00 % NaCl standards
- Auto temperature compensation (2 %/°C fixed) and temperature calibration from 5 to 40 °C
- IP67 dust- and waterproof, replaceable sensor, ~400 h battery life on 2 × CR2032, packed in handy carrying case
Overview
The only pocket meter that directly measures conductivity in 0.12ml sample and converts it into salinity or salt concentration. No need for a beaker to calibrate the meter or measure a sample. Just place few drops of the standard or sample onto the sensor. This procedure saves you time and prevents wasting your precious sample.
Specifications
| Model | SALT-11 |
|---|---|
| Measurement principle | 4-electrode method |
| Minimum sample volume | 0.12 mL |
| Salinity range (g/L) | 0.0 to 100.0 g/L (ppt) |
| Resolution (g/L) | 0.1 g/L |
| Salinity range (%) | 0.00 to 10.00% |
| Resolution (%) | 0.01% |
| Calibration points | Up to 2 |
| Calibration curves | NaCl and seawater |
| Accuracy | ±2% full scale (0.0 to 9.9 g/L), ±4% full scale (10.0 to 100.0 g/L) |
| Temperature display | 0 to 50.0 °C |
| Temperature calibration | 5 to 40.0 °C |
| Display | Custom monochrome digital LCD with backlight |
| Operating temperature | 5 to 40.0 °C |
| Operating humidity | 85% RH or less (no condensation) |
| Power | 2 × CR2032 batteries |
| Battery life | Approx. 400 hours continuous use without backlight |
| Material | ABS epoxy body |
| Dimensions | 164 × 29 × 20 mm (excluding projections) |
| Weight | Approx. 45 g (meter only, excluding batteries) |
| Waterproof rating | IP67 (dust and waterproof) |
| Functions | Temperature compensation (2%/°C fixed), Temperature calibration, Auto hold / auto stable measurement, Switchable backlight, Automatic power off (30 min), Low battery indicator, Replaceable sensor |
Applications
Used for direct measurement of salinity (NaCl or seawater equivalent) in food, beverages, pickling brines, sauces, soups, aquaculture water, seawater and soil extracts. Widely adopted in food quality control (salt content), aquaculture and aquarium management, and field salinity screening.
Setup guide
- Insert 2 × CR2032 batteries into the meter.
- Fill the measurement cell to the top with the 0.50 % NaCl standard (514-05), close the cover, wait about one minute and press the calibration key (CAL).
- Rinse with DI water, then fill the cell with the 5.00 % NaCl standard (514-50) and repeat.
- Rinse again, then apply 0.12 ml of the sample.
- Wait for the auto-stable indicator and read the salinity in g/L or % NaCl.
Compatibility
HORIBA 514-05 (0.5 % NaCl) and 514-50 (5.0 % NaCl) calibration standards; HORIBA 514-4EC-CON conditioning solution; HORIBA LAQUAtwin 4M Kit (1300083234, optional accessory); CR2032 batteries; HORIBA rinse bottle.
Precautions
Handle the flat sensor carefully — do not scratch the conductivity electrodes with sharp or abrasive objects. Always rinse the sensor with DI water after measuring corrosive or high-ionic-strength samples. Replace the sensor cap when not in use to prevent contamination.
FAQ
What does the LAQUAtwin SALT-11 actually measure, and is it the same as sodium?
The SALT-11 measures salinity as NaCl using a four-electrode conductivity method, not sodium ions directly. It reads 0.0 to 100.0 g/L (0.00 to 10.00 percent) in about 0.12 mL of sample and offers NaCl and seawater curves. It suits brines, soups and sauces; use the NA-11 when you need sodium itself.
How do I calibrate the SALT-11 and which standards does it use?
The SALT-11 uses a two-point calibration with 5 ppt (0.5 percent) and 50 ppt (5 percent) NaCl standards. Fill the measurement cell to the top without bubbles, close the cover, wait about one minute until the reading stabilises, then press CAL. Rinse with pure water before the second standard.
Does the SALT-11 sensor need conditioning before first use?
Yes. Before first use, and when response or repeatability is poor, place some drops of conditioning solution on the measurement part, wait 20 minutes, then wash with pure water and calibrate. A bottle of conditioning solution comes with the meter, and HORIBA lists replacement conditioning solution as 514-4EC-CON.
How accurate is the SALT-11 and what samples should I avoid?
Accuracy is plus or minus 2 percent of full scale from 0.0 to 9.9 g/L and plus or minus 4 percent of full scale from 10.0 to 100.0 g/L, with fixed 2 percent per degree temperature compensation. Avoid organic solvents, oils, adhesives, cement, alcohols, surfactants and strong acids or alkalis, which can damage the sensor.
How do I store the SALT-11 and how long do the batteries last?
Rinse the sensor with pure water after use and store the meter with the sensor attached in its case at 5 to 40 degrees C, away from sunlight, never immersed in water. Two CR2032 batteries give about 400 hours of continuous use with the backlight off. The sensor is a consumable and can be replaced.