K Potassium Standard solution 150ppm 6 x 14 ml | Each (3200457720)Price on request
SKU: 3200457720Brand: HORIBA

K Potassium Standard solution 150ppm 6 x 14 ml | Each (3200457720)

Key features

  • HORIBA Y-series Potassium (K⁺) standard solution at 150 ppm (mg/L)
  • Supplied as 6 × 14 ml dropper bottles— recap after use and never return used solution to the bottle
  • Used as the low calibration point for the matching LAQUAtwin ion meter
  • Convenient 14 ml size fits directly onto the LAQUAtwin flat sensor for one-drop calibration
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

HORIBA Y031L Potassium (K⁺) Standard Solution (150 ppm / mg/L) is supplied as a pack of six 14 ml dropper bottles. It is used as the low calibration point for the LAQUAtwin K-11 ion pocket meter only. Recap each bottle after use and never return used solution to the bottle.

Specifications

Article number3200457720
HORIBA modelY031L
ParameterPotassium (K⁺)
Concentration150 ppm (mg/L)
Pack6 × 14 ml dropper bottles
CompatibilityLAQUAtwin K-11
Storage5-30 °C, tightly closed, away from direct sunlight
Shelf lifeSee bottle label
BrandHORIBA Advanced Techno, Co., Ltd.

Applications

Used as the low calibration point for the LAQUAtwin K-11 ion pocket meter. The 150 ppm and 2000 ppm standards together form the two-point calibration required before measuring Potassium (K⁺) in plant sap, soil extracts, irrigation water, food, beverage or biological samples.

Setup guide

  1. Open one 14 ml bottle of the low (150 ppm) standard.
  2. Apply a few drops directly onto the LAQUAtwin flat sensor (or onto sampling sheet B for 0.05 ml volumes).
  3. Press the calibration key — the meter will auto-recognise the 150 ppm standard.
  4. After calibration, rinse the sensor thoroughly with DI water.
  5. Repeat with the complementary 2000 ppm high standard to complete the two-point calibration, then proceed with sample measurement.

Compatibility

HORIBA LAQUAtwin K-11 pocket meter; HORIBA LAQUAtwin 4M Kit (1300083234); HORIBA LAQUAtwin sampling sheet B (Y046); HORIBA 60 ml rinse bottle.

Precautions

For calibration use only; do not ingest. Avoid contact with skin and eyes. Recap each bottle immediately after dispensing to prevent evaporation and contamination. Do not return used standard to the bottle. Dispose of expired standard according to local regulations.

FAQ

What is the Potassium 150 ppm standard used for on the LAQUAtwin K-11?

It is the low point of the two-point calibration of the LAQUAtwin K-11. The meter recognises the 150 ppm and 2000 ppm potassium standards automatically by default, so calibrate with both.

How do I calibrate the LAQUAtwin K-11 with the 150 ppm potassium standard?

Open the sensor cover, rinse the flat sensor with DI or tap water, dry it with a tissue and cover the whole sensor with a few drops of the 150 ppm standard. Close the cover, press CAL and confirm the value shown, then repeat for the 2000 ppm standard. Rinsing the sensor with a little of the next standard first improves accuracy.

Why does the K-11 need both 150 ppm and 2000 ppm potassium standards?

A two-point calibration fits the sensor slope across the working range. The two standards are about ten-fold apart, so they bracket typical samples. HORIBA advises calibrating with solutions close to the expected sample value, and the high point can be changed between 160 and 9990 ppm if you use a different standard.

What does 150 ppm potassium correspond to in mmol/L on the K-11?

150 ppm of potassium is about 3.8 mmol/L, calculated from the atomic mass of 39.1. The K-11 range is specified as 0.1 to 250 mmol/L as well as 4 to 9900 ppm, and 150 ppm is the default low calibration point. The low point can be changed from 5 to 990 ppm in the meter settings.

How should I store and handle the 150 ppm potassium standard bottles?

Keep each bottle tightly closed after dispensing and store it away from direct sunlight and heat, following any storage and expiry information printed on the label. Apply only the drops you need and never return used solution to the bottle, so the standard left in the bottle stays uncontaminated.