Hipette 8 Channel Adjustable 10-100ul (7030603008)Price on request
SKU: 7030603008Brand: DLab Scientific

Hipette 8 Channel Adjustable 10-100ul (7030603008)

Key features

  • 8-channel pipette for high-throughput microplate filling (96-well plate in 12 dispenses)
  • 10-100 μL range — common ELISA / PCR setup volume
  • Lightweight ergonomic body; color-coded for channel identification
  • Digital display; calibration adjusted with the DLab calibration tool
  • Compatible with standard 200 μL tips
PricePrice on requestIn stock
WhatsApp

Request a quote

Overview

The DLab Hipette 8-channel adjustable mechanical pipette covers the 10-100 μL range across 8 channels simultaneously, designed for high-throughput microplate workflows. The 8-channel design allows filling a 96-well plate in 12 strokes (instead of 96 with single-channel).

Specifications

ModelHipette 8-channel (Cat. No. 7030603008)
BrandDLab Scientific (Beijing)
Type8-channel, adjustable volume, air displacement, color-coded
Volume Range10-100 μL per channel
Increment (Step)1 μL
Accuracy (±%)±3.0% at 10 μL; ±1.5% at 50 μL; ±1.0% at 100 μL
Precision (CV%)≤1.5% at 10 μL; ≤0.6% at 50 μL; ≤0.4% at 100 μL
Tip Cone ColorYellow (8x, fits standard 200 μL tips)
DisplayDigital mechanical counter
CalibrationAdjusted with the DLab calibration tool; refer to ISO 8655
AutoclavabilityLower shaft assembly only (121°C)
Weight~120 g
Length220 mm
StandardISO 8655; CE-marked

Applications

Used in molecular biology (qPCR setup, microplate ELISA, cell-based assays), drug discovery (HTS screening), and any microplate-based workflow requiring repeat dispensing across 8 wells. Common in 96-well plate-based immunoassays and enzyme activity assays.

Setup guide

  1. Set volume via thumb wheel; verify on display.
  2. Mount 8 tips firmly — press against tip tray for even seating.
  3. Pre-wet by aspirating/dispensing 3 times.
  4. Press plunger to first stop; immerse tips just below the surface, no deeper than 2 mm in trough/reservoir.
  5. Aspirate; dispense into microplate; press to second stop to blow out.

Compatibility

Compatible with universal 200 μL pipette tips (8 required). Optional: replacement seal kit; calibration tool; multichannel pipette stand.

Precautions

Apply tips firmly and evenly — uneven tip mounting causes channel-to-channel variation. Do not invert pipette with liquid in tips. Pre-wet all 8 tips 3 times before measurement. Avoid immersing tips deeper than 2 mm.

Documents

FAQ

What tips fit the Hipette 8-channel 10-100 uL pipette?

The yellow tip cones take standard 200 uL tips, and 8 tips are used per aspiration. Press the cones firmly and evenly onto a tip tray so all channels seat the same; uneven seating is the usual cause of channel-to-channel variation.

How many dispenses does it take to fill a 96-well plate with the Hipette 8-channel?

The 8 channels fill one column of 8 wells per dispense, so a full 96-well plate takes 12 dispenses, one per column. Pour the liquid into a reservoir trough so all 8 tips aspirate evenly, and set a volume within the 10-100 uL range.

How should I pre-wet and use the Hipette 8-channel for accurate dispensing?

Mount 8 tips firmly and evenly, pre-wet them by aspirating and dispensing 2-3 times, then press to the first stop, immerse the tips only just below the surface (DLab's guide gives 2 mm maximum for 10-100 uL), release slowly to aspirate and dispense into the plate, pressing to the second stop to blow out. Do not lay or invert the pipette with liquid in the tips.

What accuracy and precision does the Hipette 8-channel 10-100 uL specify?

The specification table lists accuracy of +-3.0% at 10 uL, +-1.5% at 50 uL and +-1.0% at 100 uL, with CV of 1.5%, 0.6% and 0.4% at the same volumes. DLab states its specifications are guidelines and user calibration should refer to ISO 8655.