SCT-ORP-MAXI ORP maxi controller (SCT-108.005.12)
Key features
- Microprocessor-based for stable and accurate readings
- Large, clear blue-white backlit 3 1/2 digit LCD display
- Isolated 4-20 mA current output
- Dual-contact output with high and low limits (normally open and normally closed)
- Low power consumption and low internal heat
- Supplied with an ORP electrode and 10 m cable
Overview
The SCT ORP (oxidation reduction potential) maxi controller is a microprocessor-based panel-mounted instrument for online analysis, monitoring and control of ORP in water-related applications. It measures -1000 to +1000 mV on a large backlit LCD, with dual high and low limit relay contacts and an isolated 4-20 mA output. It is supplied with an ORP electrode on a 10 m cable.
Specifications
| Product no | SCT-108.005.12 |
|---|---|
| Model no | SCT-ORP-MAXI |
| Measurement range | -1000~+1000 mV |
| Accuracy | ±2 mV |
| Display mode | 3½ bit LCD with backlight |
| Output current | Isolated active output of 4~20 mA |
| Output control | Dual-contact output with both higher and lower limits (normally open and normally closed respectively) |
| Contact capacity | 7A/250V AC (resistive load) |
| Length of the cable | 10 m |
| Screw size | 3/4" pipe thread |
| Requirements on water samples | Pressure ≤0.3 MPa; flow rate ≤3 m/s |
| Power supply | AC 220V±10%; 50Hz |
| Environmental conditions | Temperature 0~50°C; humidity ≤85%RH |
| Overall dimension | 96x96x120 mm (height x width x depth) |
| Face opening | 92x92 mm |
| Installation method | Panel-mounted |
| Ordering code | SCT-108.005.12 (SCT-ORP-MAXI), supplied with ORP sensor, fixation clamp and instruction manual |
| Accessory SCT-108.005.12A | ORP probe with 10 meter cable |
| Accessory SCT-108.002.DA | 222 mV ORP standard solution, 250 mL |
| Accessory SCT-108.002.CA | Electrode storage solution, 500 mL |
| Accessory SCT-108.005.12B | Mounting clamps |
| Weight | 1.5 kg |
Applications
Online ORP measurement, monitoring and control in water-related applications.
In the box
- SCT ORP maxi controller
- ORP sensor (electrode) with 10 m cable
- Fixation clamp
- Instruction manual
- Optional: SCT-108.005.12A ORP probe with 10 meter cable
- Optional: SCT-108.002.DA 222 mV ORP standard solution, 250 mL
- Optional: SCT-108.002.CA Electrode storage solution, 500 mL
- Optional: SCT-108.005.12B Mounting clamps
Precautions
Water sample requirements: pressure ≤0.3 MPa, flow rate ≤3 m/s. Operating environment 0-50°C, humidity ≤85%RH.
Documents
FAQ
How do I check the SCT-ORP-MAXI sensor with the 222 mV ORP standard?
Rinse the sensor and place it in the ORP standard solution, then wait a few minutes for a stable reading. The expected value depends on temperature and reference type; Zobell solution reads about 228 mV against Ag/AgCl at 25 C and decreases as temperature rises, so compare with the label value. Large deviations usually mean a dirty electrode or old standard.
Why does ORP on the SCT-ORP-MAXI not equal chlorine or oxidizer concentration?
ORP is the combined potential of every redox-active species present, so a mV reading is not a concentration. It becomes a useful control signal once you link a setpoint to a chemical test, such as DPD for chlorine, on your own water. pH, temperature and contamination of the electrode can all move the reading.
What installation limits apply to the SCT-ORP-MAXI sensor?
The sample pressure should be 0.3 MPa or less and flow 3 m/s or less. The sensor has a 3/4 inch pipe thread and a 10 m cable. The controller is panel-mounted in a 92 x 92 mm opening, in ambient 0 to 50 C and up to 85% RH.
How do the relays and 4-20 mA output of the SCT-ORP-MAXI work?
It has dual contacts with a normally open high limit and a normally closed low limit, each rated 7 A at 250 V AC for resistive loads, plus an isolated active 4-20 mA output. Use the contacts to switch an oxidizer or reducing-agent dosing pump through a contactor if needed, and the 4-20 mA signal for a PLC or recorder.
How do I clean the SCT-ORP-MAXI electrode?
Soak 10 to 15 minutes in water with a little dishwashing liquid, wipe the platinum tip with a cotton swab and rinse. For stubborn films follow the electrode maker's guidance, as some makers allow a longer soak in diluted bleach with a long clean-water rinse. Recheck in the ORP standard afterwards, and store the electrode in storage solution, not dry.