SCT-CL-MAXI chlorine maxi controller (SCT-108.005.26)AED 3,600.00
SKU: SCT-108.005.26Brand: Scichemtech (SCT)

SCT-CL-MAXI chlorine maxi controller (SCT-108.005.26)

Key features

  • Measuring range 0.00-19.99 mg/L with 0.01 mg/L resolution
  • Diaphragm polarographic sensor; no reagents required, direct measurement
  • Large, clear blue-white LCD display
  • Microprocessor based for stable, accurate readings
  • Low power consumption and low internal heat
  • Isolated 4-20 mA current output
PriceAED 3,600.00In stock
WhatsApp

Request a quote

Overview

The SCT-CL-MAXI is a microprocessor-based, panel-mounted residual chlorine controller for online analysis and monitoring in water applications. It uses a diaphragm polarographic sensor, measures directly without reagents, and provides an isolated 4-20 mA output, RS485 communication and an automatic sensor cleaning control output. It is supplied with a chlorine sensor, fixation clamp and instruction manual.

Specifications

Product noSCT-108.005.26
Model noSCT-CL-MAXI
Measuring range0.00 ~ 19.99 mg/L
Resolution0.01 mg/L
Zero point calibration range2.00 mg/L
Slope of the calibration range20% ~ 180%
Temperature compensation0 ~ 45 °C (automatic or manual)
Relay control range0.00 to 19.99
Automatic cleaning time1s ~ 99m59s
Cleaning interval1 ~ 720 min
Output signal4 ~ 20 mA
Sensor typeDiaphragm polarographic sensor
Sensor cleaningautomatic cleaning control output
Serial communicationRS485 communication interface
Environmental conditions temperature0~50 °C
Humidity≤85% RH
Overall dimension96x96x120 mm (Height x Width x Depth)
Dimension of the hole92x92 mm
Installation methodpanel-mounted
Power supplyAC 220/240 V 50/60 Hz
Ordering codeSCT-108.005.26 (SCT-CL-MAXI) - SCT chlorine controller supplied with chlorine sensor, fixation clamp and instruction manual
Weight2 kg

Applications

Inline measurement and control of residual chlorine in water-related applications.

Setup guide

  1. The electrode consists of a cathode, anode, electrolyte and a gas-permeable diaphragm over the cathode. Residual chlorine diffuses through the diaphragm to the cathode, where a polarization voltage between cathode and anode reduces it; the resulting current is proportional to the residual chlorine in the measured medium.

In the box

  • SCT-CL-MAXI chlorine controller
  • Chlorine sensor
  • Fixation clamp
  • Instruction manual

Compatibility

Includes: chlorine controller, chlorine sensor (diaphragm polarographic), fixation clamp, instruction manual. Compatible with: standard 230V chlorine dosing pumps; flow cell (required); DPD calibration kit; chart recorder.

Precautions

Chlorine sensor requires constant sample flow (30-60 L/h) for accurate measurement — install in flow cell with sample supply. Sensor membrane is fragile — handle with care, never touch with bare fingers. Calibrate weekly with DPD colorimetric reference. Always de-energize power before wiring to relay contacts.

Documents

FAQ

How does the SCT-CL-MAXI measure residual chlorine without reagents?

It uses a diaphragm polarographic sensor: chlorine diffuses through a gas-permeable diaphragm to the cathode, where it is reduced at a fixed polarization voltage, and the resulting current is proportional to the chlorine concentration. The range is 0.00 to 19.99 mg/L with 0.01 mg/L resolution, with automatic or manual temperature compensation from 0 to 45 C.

Why do pH and sample flow matter for the SCT-CL-MAXI chlorine sensor?

Amperometric chlorine sensors mainly respond to hypochlorous acid, whose share of free chlorine falls as pH rises, so a pH change can shift the reading noticeably, up to about 20% per pH unit in some cases. Keep sample pH stable and supply a constant sample through a flow cell, about 30 to 60 L/h for this sensor.

How should the SCT-CL-MAXI be calibrated and checked?

The controller offers zero calibration and a slope adjustment of 20% to 180%. Compare the displayed value with a DPD colorimetric test on the same sample at the same time and adjust the slope to match, repeating regularly, typically weekly. Handle the diaphragm with care and never touch it with bare fingers.

What do the cleaning output and RS485 on the SCT-CL-MAXI do?

The automatic cleaning output switches a connected cleaning device on a timer, with adjustable cleaning time from 1 s to 99 min 59 s and interval from 1 to 720 minutes. The controller has an RS485 communication interface; its protocol and register map are not published here, so check the manual before connecting a PLC or logger. The isolated 4-20 mA output is the standard analogue alternative.

What panel cut-out does the SCT-CL-MAXI need and what is supplied?

The 96 x 96 x 120 mm panel-mount housing fits a 92 x 92 mm opening, secured with the supplied fixation clamp. It ships with the chlorine sensor, fixation clamp and instruction manual. A flow cell with a constant sample supply is needed for the sensor and is not listed as included.