LAQUAtwin EC-22, Conductivity Pocket Meter, EC/Temp, 0-199.9mS/cm, 3-point Calibration | Each (3999960126)
Key features
- Only pocket meter that directly measures conductivity in a 0.12 ml sample with 3-point calibration and temperature display
- Auto range across 199 µS/cm, 1,999 µS/cm, 19.99 mS/cm and 199.9 mS/cm
- Auto standard recognition for 1,413 µS/cm, 12.88 mS/cm and 111.8 mS/cm
- Auto temperature compensation (2 %/°C fixed) plus user temperature calibration (5-40 °C); displays conductivity and temperature simultaneously
- IP67 dust- and waterproof, replaceable sensor, ~400 h battery life on 2 × CR2032, packed in handy carrying case with 1,413 µS/cm, 12.88 mS/cm and 111.8 mS/cm standards
Overview
The only pocket meter that directly measures conductivity in 0.12ml sample. No need for a beaker to calibrate your meter or measure a sample. Just place few drops of the standard or sample onto the sensor. This procedure saves you time and prevents wasting your precious sample.
Specifications
| Model | EC-22 |
|---|---|
| Measurement principle | 2-electrode bipolar AC |
| Minimum sample volume | 0.12 mL |
| Conductivity range 1 | 0 to 199 µS/cm, resolution 1 µS/cm |
| Conductivity range 2 | 200 to 1999 µS/cm, resolution 1 µS/cm |
| Conductivity range 3 | 2.00 to 19.99 mS/cm, resolution 0.01 mS/cm |
| Conductivity range 4 | 20.0 to 199.9 mS/cm, resolution 0.1 mS/cm |
| Calibration points | Up to 3 |
| Accuracy | ±2% full scale (0 to 19.99 mS/cm), ±5% full scale (20.0 to 199.9 mS/cm) |
| Temperature display | 0 to 50.0 °C |
| Temperature calibration | 5 to 40.0 °C |
| Display | Custom (monochrome) LCD with backlight |
| Operating temperature | 5 to 40 °C |
| Operating humidity | 85% RH or less (no condensation) |
| Power | 2 × CR2032 batteries |
| Battery life | Approx. 400 hours continuous use (without backlight) |
| Dimensions | 164 × 29 × 20 mm (excluding projections) |
| Weight | Approx. 50 g (including sensor and batteries) |
| Waterproof rating | IP67 dust/waterproof |
| Functions | Auto range, Auto calibration, Auto standard recognition, Auto temperature compensation (2%/°C fixed), Temperature calibration, Auto stable / auto hold measurement, Auto power off (30 min), Switchable backlight, Low battery indicator, Replaceable sensor |
Applications
Used for direct conductivity measurement in a single drop (≥0.12 ml) of sample with 3-point calibration and temperature display. Suited to applications requiring higher accuracy and a wider calibration span, including hydroponics, drinking water, environmental water, food brines and process water quality checks up to 199.9 mS/cm.
Setup guide
- Insert 2 × CR2032 batteries into the meter.
- Remove the sensor cap and apply a few drops of 1,413 µS/cm standard onto the flat sensor; press the calibration key — the meter auto-recognises 1,413 µS/cm.
- Rinse with DI water, then apply 12.88 mS/cm standard and repeat.
- Rinse again, then apply 111.8 mS/cm standard for the 3rd calibration point.
- Rinse, then apply 0.12 ml of the sample; wait for the auto-stable indicator and read the conductivity and temperature values.
Compatibility
HORIBA 1,413 µS/cm, 12.88 mS/cm and 111.8 mS/cm conductivity standards (14 ml or 514-22/514-23/514-24 small bottles); HORIBA 514-20 conditioning solution; HORIBA rinse bottle; CR2032 batteries; replacement EC sensor (model-specific). Compatible with the LAQUAtwin 4M Kit case as an accessory meter.
Precautions
Handle the flat sensor carefully — do not scratch the platinum-black electrode coating with sharp or abrasive objects. Do not wipe the platinum-black surface — gently blot dry. Always rinse the sensor with DI water after measuring corrosive, fouling or high-ionic-strength samples. Use HORIBA 514-20 conditioning solution if response becomes slow.
FAQ
Does the LAQUAtwin EC-22 measure TDS?
No. The EC-22 measures conductivity and temperature. TDS with selectable curves is a feature of the EC-33. The EC-22 reads 0 uS/cm to 199.9 mS/cm with auto range and shows temperature from 0 to 50.0 C.
Which calibration standards does the EC-22 recognise and how many points can I use?
The EC-22 supports up to 3 calibration points and auto-recognises 1413 uS/cm, 12.88 mS/cm and 111.8 mS/cm standards. The 111.8 mS/cm point extends calibration into the high range. Place a few drops on the flat sensor, calibrate, and rinse between standards. Condition a new sensor and calibrate with fresh standards.
How accurate is the EC-22 at high conductivity?
Accuracy is plus or minus 2 percent of full scale from 0 to 19.99 mS/cm and plus or minus 5 percent of full scale from 20.0 to 199.9 mS/cm. Resolution steps from 1 uS/cm at low range to 0.1 mS/cm above 20 mS/cm, selected automatically.
How do I maintain the EC-22 sensor and what is the replacement part?
Condition with conditioning solution for 10 to 30 minutes before first use, store the clean sensor dry, and for stubborn deposits use Cleaning Solution 220. Never use organic solvents such as acetone or ethanol. The replacement is the S070 sensor, part 3200459672, which is not compatible with the newer 4EC series.
How does the EC-22 compensate for temperature?
It applies automatic temperature compensation with a fixed 2 percent per degree C factor and displays temperature from 0 to 50.0 C. A temperature calibration function is also available. The meter operates between 5 and 40 C with humidity of 85 percent or less without condensation.